SUMIPAX DI-NO2 SG-05 series

Exhibits high resolution for recognizing Positional isomer

For SUMIPAX DI-NO2 SG, 3,5-dinitrophenyl group is bound to silica gel by covalent bond via a spacer, useful for recognizing positional isomer of aromatic compounds and substances with unsaturated bonds.
It is usable for both normal phase and reversed phase systems.

SUMIPAX DI-NO2 SG-05 series

Product specification

Product name
SUMIPAX
Separation mode Functional group Particle size
(μm)
Pore size
(Å)
Main target compounds
DI-NO2 SG-05 series Normal and reversed phase 3,5-dinitrophenyl 5 120 positional isomer of aromatic compounds

Applications

Positional isomer (terphenyl)

SUMIPAX DI-NO2 SG-05 series

Particle size
(μm)
Product name
SUMIPAX
Internal diameter
(mm)
Length
(mm)
Product code
5 DINO2 SG-05-4625 4.6 250 ASG05-4625W
DINO2 SG-05-20025 20 250 ASG0520025W
  • Note : Please use SUMIPAX Filters for guard columns
  • Click here for SUMIPAX Filters

Contact Us for HPLC Columns

Please contact us for any information on related to HPLC columns.

Telephone
18301939375
FAX
info@jinpanbio.com

No.5 检测波长的设定


No.5 检测波长的设定

检测波长的设定常常是根据目的成分的UV吸收值来确定的,但请不要只注意追求吸收值的最大化。对于某些物质,有时即使波长仅改变2~3nm,吸光系数也会有很大变化。
由于HPLC系统间存在设备差异,因此即使是想设置成同一波长,在实际测定中也会存在少量偏差。为了避免因2~3nm波长的差异产生的影响,其关键是先确认所设定波长的附近波长对吸光系数影响程度不大。

No.8 使用短柱时存在的问题


No.8 使用短柱时存在的问题

No.8 使用短柱时存在的问题

为了缩短分析时间,当我们将色谱柱从常规型号(150*4.6mml.D.)更换成短柱(75*4.6mml.D.)时,可能会出现理论塔板数降低、峰形拖尾的现象(图A)。此类问题可通过更换进样器得到了解决(图B),这是为什么呢?

这是由于更换前后进样器的差异引起样品扩散程度的不同造成的。一般柱外扩散的影响对保留时间短的峰影响偏大。基于此,对于常规色谱柱基本没有多大影响的扩散作用,在使用短柱时对色谱柱的峰形可能会有较大影响。在化合物的分离过程中,当出现理论塔板数偏低,峰形不佳等问题时虽然常常被认为是色谱柱引起的,但实际上还有可能是由于系统内的死体积等硬件方面引起的。鉴于短柱分离时样品扩散常常被忽略,在使用短柱时推荐使用细短配管和小进样器等可以减少扩散的部件。